PCS.smart+ AMS
Error 03
• Microcontroller
fault / MCP2
Error 04
• EEPROM error
Error 22
• Moisture de-
tector 1 cable
break
Error 32
• Moisture de-
tector 2 cable
break
Error 40
• General error
temperature
sensor 1 (block
temperature)
Error 41
• Low temperat-
ure / short-cir-
cuit temperat-
ure sensor 1
Error 42
• Excess temper-
ature / short-
circuit temper-
ature sensor 1
Error 43
• Measurement
fluctuation
temperature
sensor 1
Error 50
• General error
temperature
sensor 2
(heated line)
Error 51
• Low temperat-
ure / short-cir-
cuit temperat-
ure sensor 2
Error 52
• Excess temper-
ature / short-
circuit temper-
ature sensor 2
Error 53
• Measurement
fluctuation
temperature
sensor 2
Status text
Possible cause
H2o.1
• Moisture alarm
moisture de-
tector 1
init
• Heated line ini-
tialisation/
heat-up phase
PuMP
• Pumps deactiv-
ated
BX464007 ◦ 10/2021
• Disconnect from
power for ap-
prox. 5 s
• Contact service
• Contact service
• Contact service
• Check moisture
detector line
• Check moisture
detector
• Check moisture
detector line
• Check moisture
detector
• Possible sensor
failure
• Check temperat-
ure sensor con-
nection
(Flashing)
• Check temperat-
ure sensor con-
nection
• Check temperat-
ure sensor con-
nection
• Possible sensor
failure
• Check temperat-
ure sensor con-
nection
• Check temperat-
ure sensor con-
nection
• Check temperat-
ure sensor con-
nection
Action
• Dry
• Check condens-
ate trap
• Wait
• Reactive pumps
via menu
dt
Active Delta T con-
trol only: The block
temperature is out-
side the defined
temperature range.
• Cooler is still in
the "break-in
phase"
• Fluctuating
ambient tem-
perature
• Insufficient
cooling capa-
city
Manu
• Gas pump was
manually activ-
ated for 30
seconds
Mode
• Heated line
Addon
with Pt100 de-
tected, regu-
lator inactive
• Excess/low
temperature
Bühler Technologies GmbH
• Wait to see if the
target temperat-
ure will be
reached
• Check the ambi-
ent temperat-
ure / power sup-
plied
• Process-related:
Adjust alarm
limits
• Not required
• Activate regu-
lator in the
menu
• see chapter
"Troubleshoot-
ing"
19