DANGER
Potentially explosive atmosphere
Explosion hazard if used in hazardous areas.
The device is not suitable for operation in hazardous areas
with potentially explosive atmospheres.
Do not expose the device to combustible or explosive gas mix-
tures.
CAUTION
Hot surface
Burning hazard
Let the device cool down before maintaining.
CAUTION
Tilting risk
Damage of the device
Secure the device against any sudden translocation during
maintenance.
3 Transport and storage
The portable gas conditioning system PCS.smart+ AMS must
be stored and transport in the original case. Operation
without case is prohibited. Only transport the device in its
designated position (upright). If this cannot be ensured due to
logistics (e.g. transport via shippers or air transport) it's im-
portant to completely drain condensate from the gas condi-
tioning system, as condensate could otherwise flow back into
the gas lines. Purge the gas path with ambient air to allow
acidic condensate to escape.
DANGER
Toxic, corrosive condensate
Protect yourself from toxic, corrosive condensate when per-
forming any type of work. Wear appropriate protective equip-
ment.
The system must be flushed with dry air to ensure the driest
possible condensate path.
The equipment must be protected from moisture and heat
when not in use. They must be stored in a covered, dry and
dust-free room at a temperature between -20 °C and 60 °C.
Outdoor storage is prohibited. On principle, the operator
must meet all applicable standards with respect to preventing
damage to the equipment due to lightning, which could result
in equipment damage.
Storage spaces must be free from ozone generating equip-
ment, e.g. fluorescent lighting, mercury vapour lamps, high
voltage electrical equipment.
4 Setup and connection
Check the device for damage prior to installation. This/these
could be a damaged housing or add-on components visible
from the outside, such as filter and flow meter. Never use
equipment with obvious damage.
12
Bühler Technologies GmbH
Please note, opening and equipping the case may change the
centre of gravity and the case will need to be braced. Add-ons
such as heated line or container should be arranged and set
up so the case cannot tip over.
CAUTION
Damage/health hazard due to heater leakage
Check the device for obvious leaks before every use and at
regular intervals. Any leaks must be repaired prior to using
the device. In addition to gas leaks, leaking fluids in particular
can post an electrical and health hazard.
4.1 Installation site requirements
Be sure the equipment is positioned upright and vertical on a
level, solid surface. Also be sure to observe the approved ambi-
ent temperature.
Please be sure to secure the device to prevent tipping and
damage from the outside.
Do not obstruct the convection of the cooler. The cooler's am-
bient air intake must not be obstructed.
CAUTION
Damage to the device
Protect the equipment against dust, falling objects and ex-
ternal impacts.
Stroke of lightning
Outdoor installation is forbidden. As a matter of principle, the
operator must regard all applicable standards according pre-
vention of damage due to lightning, which may otherwise
damage the device.
4.2 Connecting a gas probe
The sample gas probe connects to the DN 6 hose connection
on the device marked IN.
The following devices (e.g. analyser) must be connected to the
DN 6 hose connection marked OUT with a suitable hose.
4.3 Electrical connections
WARNING
Hazardous electrical voltage
The device must be installed by trained staff only.
CAUTION
Wrong mains voltage
Wrong mains voltage may damage the device.
Regard the correct mains voltage as given on the type plate.
NOTICE
Separator
The supply cable also serves as separator. Ensure the supply
cable
• is clearly identifiable
• easy to reach.
The cable must not be longer than 3 m.