1 General

In a professional setting, do not use the support strap or the stor-
age pouch, because neither of these items can be disinfected or
sterilised.
Ensure that the system is processed ...
NOTICE
Ensure that the system is processed for re-use only with the
application of adequately validated device- and product-specific
procedures, and that the validated parameters are maintained
during each cycle. The equipment used must be maintained
regularly.
Follow the schedule defined below for hygienic re-use:
– Single patient use:
Clean the PARI O-PEP after every application.
Disinfect it at least once a month.
– Multiple patient use:
Clean, disinfect and sterilise the PARI O-PEP before every
change in patient.
The PARI O-PEP must be processed for re-use with the applica-
tion of suitable, validated methods in such manner that the suc-
cess of these methods can be monitored reliably and the safety
and health of patients is not threatened.
Ensure that the components are allowed ...
CAUTION
Ensure that the components are allowed to dry sufficiently after
each processing step. Condensation or residual wetness can
present an increased risk through the growth of bacteria.