5.3 Manual disinfection

For manual disinfection, select a suitable instrument disinfectant
that has been proven to be effective and compatible with the
cleaning agent used.
EQUIPMENT
– Disinfectant
– Water
PARI product components are not ...
CAUTION
PARI product components are not suitable for treatment with
instrument disinfectants based on quaternary ammonium
compounds. Quaternary ammonium compounds tend to
become concentrated in plastic materials, which can cause
incompatibility reactions in patients.
PROCEDURE
• Place all the parts in the disinfection solution.
• Allow the solution to take effect as indicated in the instructions
for use of the disinfectant.
• Thoroughly rinse the parts under hygienically tested water
and leave them to dry completely on a clean, dry and
absorbent surface.
VALIDATION
The method has been validated using:
®
– Korsolex
Basic 4% (Bode)
– Treatment time: 15 minutes
- 19 -