2.4 Closing the terminal
Closing IND221
➜ Mount the cover and tighten the 4 screws.
Closing IND226
➜ Mount the cover and press on the bottom housing until the spring clips engage.

2.5 Connecting the power supply

CAUTION!
Before connecting the weighing terminal to the power supply, check that the
voltage value printed on the rating plate corresponds with the local power
voltage.
▲ Do not connect the weighing terminal if the voltage value printed on the
➜ Plug the power plug into the socket.
After connection, the weighing terminal performs a self-test.
When the zero display appears, the weighing terminal is ready for operation.
➜ Adjust the weighing platform in order to obtain the highest possible resolution (see
Service Manual).

2.6 Inserting/replacing battery

WARNING!
Risk of explosion due to wrong battery type or improper connection!
▲ Only use original METTLER TOLEDO batteries.
▲ Make sure that the battery is connected to the terminal properly.
1. Open the battery case on the bottom of the terminal.
2. Undo the old rechargeable battery from the velcro strip, and disconnect plug if necessary.
3. Connect the (new) rechargeable battery to the terminal, and insert the rechargeable battery
in the battery case.
4. Close the battery case.
rating plate is different to the local voltage.