Clearing the tare
➜ Press
.
C
The Net indicator goes out, the gross weight appears in the display.
• If automatic clearing of the tare weight is set in the supervisor menu (F1.5.2=On), the
tare weight is cleared automatically as soon as the weighing platform is unloaded to zero.
• If tare interlock is set in the supervisor menu (F1.5.3=On), the tare weight can only be
cleared when the weighing platform is unloaded to zero.
Automatic taring
This function must be activated in the supervisor menu (F1.5.1=On).
➜ Place the empty container on the weighing platform.
The weight applied on the weighing platform is automatically saved as the tare weight.
The zero display and the Net indicator appear.
Printing/transferring data
➜ Press
.
The display contents are printed out or transferred to a computer.
Note
The display contents will not be printed or transferred if the scale is in motion.
3.5 Notes on battery operation
With a fully charged new rechargeable battery, there are approx. 35 operating hours at stan-
dard usage level.
The
indicator shows the state of the rechargeable battery.
+
constantly red
+
slowly blinking, red
+
fast blinking, red
+
+
green
Approx. 10 % of full voltage
Approx. 5 % of full voltage
Less than 5 % of full voltage,
battery has to be charged immediately
Battery is charged.
Charging takes place as soon as the weighing terminal
is connected to external power.