Connecting the Weighing Platform to the
Weighing Terminal
The weighing platform PBA220 is designed for use with
analog METTLER TOLEDO weighing terminals.
1.
Insert the weighing platform cable through the cable
gland into the weighing terminal.
2.
Connect the weighing platform cable to the 7-pin
terminal strip J2 of the weighing terminal according
to the following tables.
PBA220-BB.../-B.../-BC.../
-CC60/150/300/ -QB.../-QC.../-QD...
Terminal
SIG+
SIG–
EXC+
EXC–
SEN+
SEN–
SHIELD
PBA220-A.../-QA...
Terminal
SIG+
SIG–
EXC+
EXC–
SHIELD
PBA220-CC600
Terminal
SIG+
SIG–
EXC+
EXC–
SEN+
SEN–
SHIELD
Color
red
white
green
black
blue
brown
yellow (long)
Color
red
white
green
black
yellow (long)
Color
white
red
green
black
yellow
blue
yellow (long)