6.4 Block F5 – Maintenance
Factory settings are printed in bold letters.
F5.1 – Calibration values
F5.1.1 – Show zero-counts
F5.1.2 – Show half load weight value
F5.1.3 – Show half load counts
F5.1.4 – Show full load weight value
F5.1.5 – Show full load weight counts
F5.2– Keypad test
The terminal shows PrESS.
T
➜ Press
.
F
C
➜ Press
to exit keypad test.
F5.3 – Display test
All display segments light up.
F5.4 – Display internal resolution
F5.5 – COM1 test
To do so, the terminal must be connected to a computer.
F5.6 – Print setup
Print all setup parameters
F5.10 – General reset
Reset all parameters of groups F1 to F4 to factory setting, except for calibration values. In OIML
and NTEP mode the Geo value is not reset.
30