F1.4.1 – Automatic zero setting
Automatic zero setting disabled
OFF
Automatic zero setting within +/–0.5 d
0.5 d
Automatic zero setting within +/–1.0 d (not available in OIML mode)
1 d
Automatic zero setting within +/–3 d
3 d
F1.4.2 – Power up zero
Power up zero disabled
OFF
Power up zero within +/–2 %
2
Power up zero within +/–10 %
10
Power up zero within +/–20 %
20
F1.4.3 – Pushbutton zero
Pushbutton zero disabled
0
Pushbutton zero with
2
+/–2 % zero setting range
Pushbutton zero with
10
+/–10 % zero setting range
Pushbutton zero with
20
+/–20 % zero setting range
F1.5.1 – Automatic taring
Automatic taring enabled
On
Automatic taring disabled
OFF
F1.5.2 – Auto clear tare
Clearing tare automatically enabled
On
Clearing tare automatically disabled
OFF
F1.5.3 – Tare Interlock
The weighing platform must be unloaded to zero before the tare weight can be
On
cleared.
Function disabled
OFF
(not available in NTEP mode)
(not available in OIML mode)
(not available in OIML and NTEP modes)
(not available in OIML and NTEP modes)
(not available in OIML and NTEP modes)