Mettler Toledo IND221 Mode D'emploi page 20

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

F2.2.2 – Setting the target weight
WEIGHt By weighing in an actual sample weight
MAnUAL By numeric entry
F2.2.3 – Upper tolerances
After selecting the parameter, the currently set tolerance value is displayed.
If F2.2.1 = Chech is set:
Upper tolerance = target value + displayed tolerance value
If F2.2.1 = CLASS is set:
Internally the terminal calculates 3 tolerances.
Upper tolerance 1 = target value + displayed tolerance value
Upper tolerance 2 = target value + 2 x displayed tolerance value
Upper tolerance 3 = target value + 3 x displayed tolerance value
1. If necessary, use the
2. Change tolerance value using the
Factory setting
Possible settings
F2.2.4 – Lower tolerances
After selecting the parameter, the currently set tolerance value is displayed.
If F2.2.1 = Chech is set:
Lower tolerance = target value – displayed tolerance value
If F2.2.1 = CLASS is set:
Internally the terminal calculates 3 tolerances.
Lower tolerance 1 = target value – displayed tolerance value
Lower tolerance 2 = target value – 2 x displayed tolerance value
Lower tolerance 3 = target value – 3 x displayed tolerance value
1. Use the
key to activate editing if necessary.
F
2. Change tolerance using the
Factory setting
Possible settings
key to activate editing.
F
upper tolerance value = 10 d
0 ... full load
,
lower tolerance value = 10 d
0 ... full load
T
,
and
F
T
and
keys.
F
20
keys.

Publicité