Cleaning and Disinfection

WARNING
To prevent cross-contamination,
always follow the disinfection instructions
in this IFU.
WARNING
To avoid material damage and injury,
clean and disinfect according to this IFU:
EN
• No other chemicals are allowed.
• Never clean with chlorine (except for
the wipeable sling).
• Chlorine will deteriorate the surface of
the material (except for the wipeable
sling).
WARNING
To avoid injury, always remove the sling
before disinfecting the lift.
WARNING
To avoid injury, never wash a Flites. A
Flites is only intended for "single patient
use".
All
Flites, should be cleaned when they are soiled or
stained and between patients.

Flites - Do NOT wash

The Flites is for single patient use.
Do NOT clean, wash, disinfect, wipe or sterilize
the Flites. If the Flites has been subjected to any
such treatment, it shall be discarded.
The Flites
symbol. (See Fig. 13 )
If the Flites
symbol will appear. (See Fig. 14 )
Fig. 13
Wipeable Sling - Cleaning
Instruction (6 steps)
1. Detach the sling from the lift.
2. If needed, rinse to remove any deposits or
visible dirt from the sling.
slings, except the
Fig. 14
10
3. Wipe all parts of the sling with one of the
cleaning agents listed under "Wipeable Sling -
4. Allow contact time according to the cleaning
agent manufacturer's instructions for use.
5. If stated in the cleaning agent manufacturer's
instructions for use, rinse to remove chemical
residue.
Air dry at room temperature.
Do NOT
machine wash

Wipeable Sling - Cleaning Agents

The sling can be cleaned using:
Ethanol based solution, up to 70%
Isopropanol based solution, up to 70%
Hydrogen peroxide based solution, up to 1.5%
Quaternary ammonium based solution,
up to 0.1%
Chlorine based solution, up to 10 000 ppm
Machine Washable Slings - Cleaning
Instruction (3 steps)
7. Detach the sling from the lift.
8. Close all buckles and hook and loop attach-
ments on the slings and belts prior to washing.
9.
please wash according to local hygiene
legislation. If allowed according to the product
label, tumble dry in low temperature, max.
Do NOT
wash with other items that have
rough surfaces or sharp objects
use any mechanical pressure, pressing
or rolling during the washing and drying
process
use bleach
use gas sterilization
use autoclave
dry clean
steam
iron
Machine Washable Slings - Cleaning
Agents
detergent without optical brightener.
No other chemicals are allowed e.g. chlorine,
softener, iodine based disinfectants, bromine
and ozone.
Flites or