Care and Cleaning

EN
The following processes is recommended, but should be adapted to comply with local
institution protocols. If you are uncertain, you should seek advice from your local Infection
Control Specialist.
The AtmosAir MRS should be routinely decontaminated between patients and at regular
intervals while in use.

Fire Barrier

Care should be taken with regards to the re barrier when the cover is removed. It is
recommended that Personal Protective Equipment (PPE) is used during handling of the
mattress with the re barrier exposed to avoid contact with loose particles. To ensure
optimal product performance, avoid over-handling of the re barrier and handle with care.
Fire barrier should be replaced if soiled or damaged.

Cover Cleaning Options

Sewn AtmosAir MRS (Top cover can not be removed from base)
1.
Remove or push bed linens to center of mattress.
2.
Wipe and rinse any soiling from the mattress surface and base.
Use 1000 ppm chlorine or 70% alcohol.
3.
Rinse with a wet towel to remove chemicals.
4.
After wipe down, dry surface with towel.
5.
Ensure bed linens are re tted and not wrinkled under patient.
6.
Clean pump and tubes (if needed) by wiping with a damp cloth.
RF Welded AtmosAir MRS (Detachable top cover)
1.
Push bed linens to center of mattress to wipe down, remove to launder top cover
or wipedown as described below.
2.
Wipe and rinse any soiling from the mattress surface and base.
Use 1000 ppm chlorine or 70% alcohol.
3.
Rinse with a wet towel to remove chemicals.
4.
After wipe down, dry surface with towel.
5.
Ensure bed linens are re tted and not wrinkled under patient.
8
Do not use Phenol-based solutions or abrasive componds or pads on
the cover during decontamination process as these will damage the
surface coating. Do not boil or autoclave the cover.
Covers should be wiped down immediately after it has been exposed
The sewn AtmosAir MRS cover is cleaned by wipe down methods only.
Do not launder as damage to the base may occur.
The RF Welded AtmosAir MRS has a detachable top cover that can be
laundered. However the base is cleaned by wipe down methods only.
Do not launder base as damage may occur.