Preparations
Before First Use (6 steps)
WARNING
To avoid injury, always make sure to
inspect the equipment prior to use. If any
part is missing or damaged - Do NOT use
the sling.
1. Check all parts of the sling, see section "Parts
Designation" on page 4. If any part is
missing or damaged - Do NOT use the sling.
2. Read this IFU thoroughly.
3. Check that the sling is clean.
4. Choose a designated area where the IFU
should be kept and easily accessible at all
times.
5. Make sure to have a rescue plan ready in
case of a patient emergency.
For questions, contact your local Arjo
representative for support.
NOTE
To keep track of the service life of the
wipeable sling, document the serial number
NOTE
Arjo recommends the name of the patient is
written on the sling care label to avoid cross
contamination between patients.
Before Every Use (8 steps)
1. Patient assessment must always determine
the methods used in everyday care, see
"Intended Use" on page 3.
WARNING
To avoid the patient from falling, make sure
to select the correct sling size according
to the IFU.
2. Check the allowed combinations against the
sling and lift, see "Allowed Combinations" on
page 7.
WARNING
To avoid injury, always make sure to
inspect the equipment prior to use. If any
part is missing or damaged - Do NOT use
the sling.
3. Check all parts of the sling, see "Parts
Designation" on page 4. If any part is
missing or damaged - Do NOT use the sling.
4. Check for:
•
fraying
•
loose stitching/delamination
•
cracks
•
tears
•
fabric holes
•
soiled fabric
•
damaged clips/loops/buckles
•
unreadable or damaged label/print
WARNING
To prevent cross-contamination, always
follow the disinfection instructions
in this IFU.
5. Check that the sling is clean. If the sling is
not clean see "Cleaning and Disinfection" on
page 10.
Always check the washing symbol on the
Flites label. If the Flites has been washed,
throw the Flites away. See "Flites - Do NOT
wash" on page 10.
7.
If this symbol appears, do NOT use
the Flites. The Flites label indicates it
has been washed.
8. Before using the sling, read the lift IFU for
transfer, transport and bathing.
After every use
If the patient is leaving the hospital, throw away
Flites.
5