Care and Cleaning
EN
Control Specialist.
AtmosAir Stretcher MRS should be routinely decontaminated between patients and at
regular intervals while in use.
Do not use Phenol-based solutions or abrasive componds or pads on
the cover during decontamination process as these will damage the
surface coating. Do not boil or autoclave the cover.
Fire Barrier
Cover Cleaning Options
Sewn AtmosAir Stretcher MRS (Top cover can not be removed from base)
1.
Remove or push bed linens to center of mattress.
The sewn AtmosAir Stretcher MRS cover is cleaned by wipe down
methods only. Do not launder as damage to the base may occur.
2.
Use 1000 ppm chlorine or 70% alcohol.
3.
Rinse with a wet towel to remove chemicals.
4.
5.
6.
RF Welded AtmosAir Stretcher MRS (Detachable top cover)
1.
or wipe down as described below.
The RF Welded AtmosAir Stretcher MRS has a detachable top cover
that can be laundered. However, the base is cleaned by wipe down
methods only. Do not launder base as damage may occur.
2.
Use 1000 ppm chlorine or 70% alcohol.
3.
Rinse with a wet towel to remove chemicals.
4.
5.
6.
8
Covers should be wiped down immediately after it has been exposed