3. OPERATIONAL, TRANSPORT AND STORAGE
SAFETY REQUIREMENTS
KEEP THE PRODUCT AND ITS ASSOCIATED EQUIPMENT AWAY FROM:
• Heat sources (fi re, cigarettes,...)
• Flammable materials
• Oil or grease (take a great care in the use of hand creams)
• Water
• Dust
The product and its associated equipment must be prevented from tipping over, turning over
or falling.
Always maintain oxygen cleanliness standards.
Only use the product and its associated equipment in well ventilated area.
Before initial use the product should be kept in its original packaging. GCE recommends the
use of original packaging if the product is withdrawn from operation (for transport, storage).
The product provider or user must ensure that during the product life cycle, transport and
storage environmental conditions are in accordance with the above "Safety instruction" section
requirements, in order to maintain product integrity and cleanliness. Statutory laws, rules and
regulations for medical gases, accident prevention and environmental protection must be
observed.
OPERATING CONDITIONS
+10* / +60 °C
10 / 100 %
600 / 1200 mbar
It's very important to assure access to this device only for qualifi ed person – possibility of
misuse.
4. PERSONNEL INSTRUCTIONS
The Medical Devices Directive 93/42/EEC states that product provider must ensure that all
personnel handling the product are provided with the operating instructions & performance
data.
Do not use the product without properly familiarization of the product and its safe operation
as defi ned in this Instruction for use. Ensure user is aware of particular information and knowl-
edge required for the gas in use.
STORAGE AND TRANSPORT
-30 / +70 °C
10 / 100 %
600 / 1200 mbar
3/100
CONDITIONS