10.4.3 Check Weighing

During weighing of a sample the balance can be set to show if the weight is above or below
an upper and a lower limit. The display will use the arrows under the bar graph to show that
check weighing is operating. The bars between the arrows will indicate when the weight is
below the lower limit, between the limits or above the upper limit.
The buzzer can be set to be active when the weight is outside the limits (below the lower or
above the upper) or within the limits (above the lower and below the upper limit), or turned
off. If desired, only one limit needs to be set. If only one limit is set the other limit is
considered to be zero (lower) or the maximum (upper).
The Check Weighing function is not active when the loaded weight is less than 20 weighing
divisions. This is the minimum weight at which the indicator bars are displayed and the
buzzer sounds (if it is not set to OFF).
Check Weighing Set-up Steps:
• Press [Mode] and then the [Up] or [Down] key to select Check Weighing. When
CHECK
CHECK
CHECK
CHECK
"
"
is displayed, press the [Select] key to enter Check Weighing mode
• Press [Up] or [Down] to set the "LOW LIMIT" to "On" or "OFF"
• Press [Select] to proceed. If the "LOW LIMIT" was set to "On", the display will show
the current low limit in the unit last used in the weighing mode. You may change the
limit using the numeric entry method (see section 8.1).
• Press [Select] to proceed.
NOTE: Low limit must be 20 weighing divisions or higher. If a value below this is
selected then
value. Re-enter a valid selection.
• If the "LOW LIMIT" was set to "OFF" or the setting of the low limit is complete, then
the display will change to "HIGH LIMIT". Use [Up] and [Down] to set the "HIGH
LIMIT" to "On" or "OFF".
• Press [Select] to proceed.
• If the "HIGH LIMIT" was set to "On", the display will show the current high limit in the
last selected unit. You may change the limit using the numeric entry method (see
section 8.1).
• Press [Select] to proceed.
NOTE: High limit must be higher than the low limit already set, and must be less than
the maximum capacity of the balance. If an incorrect value is entered then
ERROR
ERROR
ERROR
ERROR
Range'
Range
or '
Range
Range
© Adam Equipment Company 2015
ERROR
ERROR
ERROR
ERROR
'
LOWW'
LOWW
LOWW
LOWW
will be displayed followed by returning to the previous
will be displayed followed by returning to the previous value.