larger number corresponds to a larger stable zone. Selectable values: 1, 2, 5
or 10 (divisions).
Recommended value for normal use: 1
ON = Auto-zero function. Selectable values: 1, 2, 5, 10 or 15 (divisions).
AV ZERO
AV ZERO
AV ZERO
AV ZERO
OFF = Auto-zero function disabled.
Recommended value for normal use: ON, 5
COMMA
SEPARAT
SEPARAT
SEPARAT
SEPARAT
DEC PT

13.6 CALIBRATION SETUP

This menu allows the Supervisor to set the calibration parameters. Press [Select] when
"CAL SET
CAL SET" is displayed to select the calibration parameters. The options for each
CAL SET
CAL SET
parameter can be scrolled through by using the [Up] or [Down] key and pressing [Select]
to confirm choices.
NO = Operator calibration is disabled.
ENABLE
ENABLE
ENABLE
ENABLE
YES = Operator calibration is enabled.
ON = Prints out Calibration report after successful calibration.
CAL REP
CAL REP
CAL REP
CAL REP
OFF = Disabled.
ON = Enabled. Select time from 1 to 24 hours.
TIM CAL
TIM CAL
TIM CAL
TIM CAL
OFF = Disabled.
ON = Enabled. Select the temperature variation from 0.2 to 4°C which when
TEM CAL
TEM CAL
TEM CAL
TEM CAL
detected will trigger automatic calibration.
OFF = Disabled.
YES = Internal calibration enabled (if fitted).
INT CAL
INT CAL
INT CAL
INT CAL
NO = External calibration enabled.
CAL MAS = Displays the set value of the internal calibration mass (if fitted) in
INT MAS
INT MAS
INT MAS
INT MAS
grams. If after verification against an external mass it is determined that the
value of the internal mass needs adjustment, e.g. due to wear, accumulation
of dirt, etc., then this value can be adjusted by +/- 100 mg. This should only
be considered by expert users if the external reference weight is definitively
accurate and an incorrect weight reading is being given after internal
calibration. Adjustment will restore the internal calibration to the correct level
of accuracy.
Press [Mode] to advance to setting of the next menu or [
weighing.

13.7 PASSCODES

To enable the security features in this balance it is necessary to set passcodes. There are
2 passcodes called Operator Passcode and Supervisor Passcode. The Operator
Passcode allows an authorised user to operate the basic weighing functions of the balance
but will not allow access to the Supervisor Menus if the Supervisor Passcode has been
set.
Note: To change or disable a Passcode it is be necessary to enter the current passcode.
To setup passcodes:
Press [Select]. Use the [Up] and [Down] keys to cycle through options until "PASSCOd
displayed. Press [Select] again to enter this section. Use [Up] and [Down] keys to select
Operator ("PC OPER
PC OPER
PC OPER") or Supervisor ("PC SV
PC OPER
© Adam Equipment Company 2015
Set separator indicator on the display to be either a decimal point
or a comma. Also applies to the serial interface for print output.
PC SVPE
PC SV
PC SV
PE") option.
PE PE
0/T] to return to normal
41
3016612482 Rev1-Aug15
PASSCOd
PASSCOd" is
PASSCOd