5 Battery Information
Charging Operations
Caution
Do not operate the unit for more than a few minutes if
the RED Low Battery indicator is illuminated. Recharge
the Battery as soon as possible.
AC/DC Adapter Charger (Cat. No. 886111)
Primarily the Battery is charged while installed within the
LCSU 4 unit by using the standard AC/DC Adapter Charger.
Power Cord
Follow the LED status on the
Control Panel on the LCSU 4 and
charge the Battery as necessary.
LED Indicator
Red LED illuminated
Yellow LED illuminated
Yellow LED flashes
Yellow LED switches off
Green LED illuminated
10
LCSU4_multi_DFU_revI.indb 10
AC/DC Adapter Charger
Green LED
Yellow LED
Red LED
Status
Battery level low
Charging
Battery almost charged
Battery is fully charged
Connected to AC/DC
External Battery Charger (Cat. No. 886112)
The Battery can be charged externally by removing it from the
LCSU 4 unit and using the External Battery Charger.
External Battery Charger
Power cord
An empty Battery must charge up to 5 hours to reach full
capacity.
Follow the LED Indicator and charge the Battery as necessary.
LED Indicator
Status
LED not illuminated
Plugs not connected
Yellow LED flashes
Charging is pending
Yellow LED illuminated
Charging
Green LED illuminated
Battery is fully charged (*)
Red LED flashes
Charge failure
* The Battery can be left on continuous charge, even if the
Green LED is illuminated. It will not harm the units.
Caution
Do not cover the Charger. When in use, it is normal that
the Charger and the Battery becomes warm.
LED Indicator
Battery
Connector
Battery