How to Operate and Adjust Suction level
1 Unwind the Patient Tube (check that no kinks will obstruct
flow).
2 Turn the unit "ON" by pressing the button (
3 Block the Patient Tube.
3
4 Set desired suction vacuum level by
turning the Vacuum Regulator.
•
Turn clockwise (+) to increase vacuum
•
Turn counter-clockwise to decrease
vacuum
5 The vacuum level will display on the 50 – 550+ mmHg
scale.
6 When the desired vacuum level is reached; unblock and
then block the Patient Tube.
Result: Unit should return to the desired level.
7 Apply necessary suction therapy. If desired, a suction tip or
catheter could be attached to the Patient Tube.
Important
If LCSU 4 does not maintain the desired suction
performance level, refer to Troubleshooting instructions
(see Chapter 7).
After Each Use
1 When finished with suctioning, let LCSU 4 run for a
moment to allow all suctioned material to flow from the
Patient Tube into the Canister.
2 Disconnect and dispose of Canister and Patient Tube.
3 Clean the exterior and any reusable parts of the LCSU 4
assembly according to instructions (see Chapter 6).
4 Perform Device Test (see Chapter 6).
5 Place Battery on charge (see Chapter 5).
LCSU4_multi_DFU_revI.indb 9
).
4
Battery Charging
Important
Use only Laerdal Battery Cat. No. 886113.
Caution
Use of other than Laerdal brand Battery may result in
errors related to the Battery status indicator, reduction of
the Battery operation time, failure to effectively operate
LCSU 4, and/or give rise to hazards to operator and/or
patient.
When in operational use
1 An empty Battery must charge up to 5 hours to reach full
capacity.
2 Battery run time: Approximately 45 minutes of continuous
operation at zero vacuum level (free flow), fully recharged.
3 Always fully charge Battery - before and after each use.
To prolong Battery lifetime it is recommended to place the
Battery on continuous charge. It will not harm the unit.
If continuous charging is not possible, charge Battery for
minimum 24 hours once a month. Completely discharging the
Battery will shorten its useful life.
When stored unused for > 3 months
•
Fully charge Battery before storing
•
Recharge every 3-6 months
Battery Test
When in operational use, the Battery should be tested every
6-12 months.
1 Start the test with a fully charged Battery.
2 Set maximum vacuum level.
3 Let the unit run for 20 minutes (free flow).
4 Block the Patient Tube.
5 If vacuum level fails to reach 550+ mmHg the Battery
should be replaced.