Cleaning And Maintenance; Cleaning; Device Test - Laerdal Compact Suction Unit LCSU 4 Mode D'emploi

Masquer les pouces Voir aussi pour Compact Suction Unit LCSU 4:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 31

Cleaning

Cautions
Disconnect the LCSU 4 from external power prior to
cleaning.
To prevent any electrical shock hazard, use minimum
liquids when cleaning. Do not immerse the LCSU 4, or
allow it to stand in water or other liquids. This can damage
the device and/or cause electrical shocks resulting in injury
to persons.
Warning
Do not pump any cleaning solution or other liquids
through the vacuum pump, i.e. through the Vacuum
Connector. This can damage the LCSU 4.
Main Cabinet
1 Disconnect from external power supply.
2 Clean Cabinet surfaces by carefully wiping with a soft cloth
or sponge with mild detergent. Use hand dishwashing liquid
or similar that is compatible with the Material Chart (see
Chapter 10).
3 Dry all surfaces using a clean cloth or paper towel.
Canisters and Patient Tubes
Dispose after use.
Important
The Canisters and Patient Tubes are all disposable items.
Do not attempt to clean or reuse these. Due to the risk
of cross contamination, all disposable items must be
replaced after each use. They are for single-patient use
only.
Vacuum Tube (for 800 ml version) and Wire Stand
Wash by immersing and scrubbing in hand dishwashing liquid
or similar.
1 Rinse thoroughly in pure water.
2 Allow to dry. Disinfect if desired.
High Efficiency Filter Kit (for 800 ml Version)
Filter cannot be cleaned or disinfected.
Replace Filter immediately if contamination or
discoloration is observed, or if it gets wet.
* If the unit is used on patients in areas where cross
contamination is an issue, it is recommended that the Filter is
replaced after each use.
LCSU4_multi_DFU_revI.indb 11

6 Cleaning and Maintenance

Carry Bags
Wipe Bags according to instructions provided above for Main
Cabinet. Do not launder.

Device Test

After each reassembly, and before returning a LCSU 4 to
operational use, the Device Test should be performed:
1 Start the test with a fully charged Battery
2 Turn the unit "ON"
3 Block the Patient Tube
4 Set Suction Vacuum level to 550+ mmHg
5 Unblock and then Block the Patient Tube again.
6 Result: Unit should return to 550+ mmHg setting.
7 Repeat procedure for 300 mmHg and 50 mmHg settings.
Cautions
Do not attempt to use any LCSU 4 that has not passed
the above test. If the LCSU 4 unit does not test
satisfactorily, recheck all parts of the assembly and test
once again. If necessary, contact Laerdal Medical or one
of its authorized distributor.
Disposal
When discarding the LCSU 4, we recommend it be discarded
according to local protocol.
This appliance is marked according to the European directive
2012/19/EU on Waste Electrical and Electronic Equipment
(WEEE). By ensuring this product is disposed of correctly,
you will help prevent potential negative consequences for the
environment and human health, which could otherwise be
caused by inappropriate waste handling of this product.
The symbol on the product, or on the documents
accompanying the product, indicates that this appliance
may not be treated as household waste. Instead it shall
be handed over to the applicable collection point for the
recycling of electrical and electronic equipment.
Disposal must be carried out in accordance with local
environment regulations for waste disposal.
For more detailed information about treatment, recovery
and recycling of this product, please contact your local city
office, your household waste disposal service or the Laerdal
representative where you purchased the product.