Operating and Maintenance Instruction FCU 8000-1 series
Then shut-off valve V2 is opened and oil begins to flow. Once the BS filling time has
elapsed, the pre-set pre-run volume is fed through.
Directly after the rinsing process, the first measurement starts.
The following message appears in the display during measurement:
Where:
M6 : measuring mode
xx : number of the current measurement
yy : progress of the measurement in %
zz : actual flow rate
qq : progress of the measurement (volume measured so far)
Once each individual measurement has been completed, the relevant result is printed out
when online printing is activated. The given particle quantity is projected for 100 ml!
If "yes" was selected in the Setup-Menu under additional measurement, you will be asked
whether an additional individual measurement (measurement volumes as for completed
individual measurements) should be carried out, once the last individual measurement
has ended.
Display :
M6: +1 test -> OK
Once the last measurement has ended or when the STOP key has been pressed, the
pressure in the chamber is released and the measurement tank drained.
The individual measurements which are to be used to produce an average overall result,
can be selected afterwards.
Display :
M6 Average value
Meas.no:123456789
The number of the individual measurements can selected using the
the relevant individual measurement can be activated (number is displayed) or
deactivated (underscore is displayed) using the
Example : Out of 5 individual measurements, the individual measurements
2, 3 and 4 are to be used to calculate an average value.
Display :
HYDAC Filtertechnik GmbH
M6: xx .Measurement
zz ml/min
Vol.: qq ml
finish
->STOP
10
M6 Average value
Meas.no:_234_
Doc.: 3109776 Ed.: 01/13.02.2002
yy %
+
-
keys for averaging.
E
keys and
Page 101