Table 3 Sensor DIAG/TEST menu (continued)
Option
Description
SENSOR DAYS
Shows the number of days that the sensor has
been in operation.
RESET SENSOR
Resets the number of days that the sensor has
DAYS
been in operation.

6.5 Error list

When an error occurs, the reading on the measurement screen flashes
and all outputs are held when specified in the controller menu. To show
the sensor errors, press the MENU key and select Sensor Diag, [Select
Sensor], Error List. A list of possible errors is shown in
Table 4 Error list for pH and ORP sensors
Error
Description
PH TOO
The measured pH is > 14
HIGH
ORP TOO
The measured ORP value
HIGH
is > 2100 mV
PH TOO LOW The measured pH is < 0
ORP TOO
The measured ORP value
LOW
is < –2100 mV
OFFSET TOO
The offset is > 9 (pH) or
HIGH
200 mV (ORP)
OFFSET TOO
The offset is < 5 (pH)
LOW
or –200 mV (ORP)
SLOPE TOO
The slope is >
HIGH
62 (pH)/1.3 (ORP)
Table
4.
Resolution
Calibrate or replace the
sensor.
Calibrate or replace the
sensor.
Follow the maintenance
procedures for the
sensor and then repeat
the calibration, or replace
the sensor.
Repeat the calibration
with a fresh buffer or
sample, or replace the
sensor.
Table 4 Error list for pH and ORP sensors (continued)
Error
Description
SLOPE TOO
The slope is <
LOW
50 (pH)/0.7 (ORP)
TEMP TOO
The measured temperature
HIGH
is >130 °C
TEMP TOO
The measured temperature
LOW
is < –10 °C
ADC
The analog to digital
FAILURE
conversion failed
A ELEC TOO
The active electrode
HIGH
impedance is > 900 MΩ
A ELEC TOO
The active electrode
LOW
impedance is < 8 MΩ
R ELEC TOO
The reference electrode
HIGH
impedance is > 900 MΩ
R ELEC TOO
The reference electrode
LOW
impedance is < 8 MΩ
SAME
The buffers for 2 point
BUFFER
buffer calibration have the
same value
SENSOR
The sensor is missing or
MISSING
disconnected
Resolution
Clean the sensor, then
repeat the calibration, or
replace the sensor.
Make sure that the
correct temperature
element is selected.
Power off and power on
the controller. Call
technical support.
The sensor is in air.
Return the sensor to the
process.
The sensor is damaged
or dirty. Call technical
support.
Buffer leaked or
evaporated. Call
technical support.
The reference electrode
is damaged. Call
technical support.
Complete the steps in
Test the pH sensor
on page 17.
Examine the wiring and
connections for the
sensor and for the
module.