Error list

Errors may occur for various reasons. The reading on the measurement
screen flashes. All outputs are held when specified in the controller
menu. To show the sensor errors, push the MENU key and select
Diagnostics, [Select Sensor], Error List. A list of possible errors is shown
in
Table
5.
Table 5 Error list for the sensor
Error
Description
SENSOR MISSING
The sensor is missing or
disconnected
Warning list
A warning does not affect the operation of menus, relays and outputs. A
warning icon flashes and a message is shown on the bottom of the
measurement screen. To show the sensor warnings, push the MENU
key and select Sensor Diag, [Select Sensor], Warning List. A list of
possible warnings is shown in
Table
Table 6 Warning list for the sensor
Warning
Description
TOTALIZER FULL
The sensor
volume/totalizer is full.
TEMP WARNING
The temperature is out of
scope.
ECHO MISSING
The echo signal is lost.
Resolution
Examine the wiring and
connections for the
sensor and for the
module.
6.
Resolution
Reset the sensor
volume/totalizer.
Replace the sensor.
Option 1—Move the
sensor closer to the
water surface.
Option 2—Adjust the
sensor mounting to
properly aim the sensor
at the water surface.
Option 3—Replace the
sensor.
Table 6 Warning list for the sensor (continued)
Warning
Description
EXCESS LEVEL
The excess level is out of
range.
REPLACE SENSOR
The sensor needs to be
replaced.
CAL DUE
The sensor calibration is
over due.
NOT CALIBRATED
The sensor needs to be
calibrated.
CAL IN PROGRESS
The calibration was
started but not
completed.
OUTPUTS ON HOLD
During calibration, the
outputs were set to hold
for a selected time.

Event list

The controller provides one event log for each sensor. The event log
stores a variety of events that occur on the devices such as configuration
changes, alarms, warning conditions, etc. A list of possible events is
shown in
Table
7. The event log can be read out in a CSV format. For
instructions on downloading the logs, see the controller user manual.
Table 7 Event list for the sensor
Event
Description
POWER ON EVENT
The power was turned on.
CHANGE IN CONFIG - float
The configuration changed-floating number
CHANGE IN CONFIG - int
The configuration changed-integer number
CHANGE IN CONFIG - text
The configuration changed-text string
Resolution
Option 1—Make sure the
gauge type is correct.
Option 2—Calibrate the
sensor.
Replace the sensor.
Calibrate the sensor.
Calibrate the sensor.
Return to calibration.
The outputs will become
active after the selected
time period.