1
Use according to specification
The centrifuge was designed only for the separation of substances or mixtures with a density of no more than 1.2
kg/dm³, particularly of samples for preparation for human medical in-vitro diagnostics using blood collection tubes
with separation gel.
The samples centrifuged with the EBA 270 can be used directly in in-vitro diagnostic tests. Therefore, the EBA 270
centrifuge as an in-vitro diagnostic accessory is in its turn an in-vitro diagnostic medical device within the meaning of
the IVD Directive 98/79/EC.
The centrifuge is only meant for this purpose.
Another use or one which goes beyond this, is considered to be non-intended. The company Andreas Hettich GmbH
& Co. KG is not liable for damage resulting from this.
Observing all information in the operating instructions and complying with the measures described therein is also a
part of the intended use.
2
Remaining risks
The device is built according to the state-of-the-art and the recognized safety regulations. If used and handled
improperly, there could be life-threatening danger to the user or third parties, or the device could be impaired or there
could be other property damage. The device is only to be used for its intended purpose and only when it is in safe
working condition.
Malfunctions which could affect safety must be corrected immediately.
3
Technical specifications
Manufacturer
Model
Type
Mains voltage ( 10%)
Mains frequency
Connected load
Current consumption
Max. capacity
Allowed density
Speed (RPM)
Force (RCF)
Kinetic energy
Obligatory inspection (BGR 500)
Ambient conditions (EN / IEC 61010-1)
Set-up site
Altitude
Ambient temperature
Humidity
Excess-voltage category
(IEC 60364-4-443)
Pollution degree
Device protection class
EMC
Emitted interference,
Interference immunity
Noise level (dependent on rotor)
Dimensions
Width
Depth
Height
Weight
AB2300DEENFRIT
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
2300
200 - 240 V 1
50 - 60 Hz
130 VA
0.7 A
Up to 2000 m above sea level
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
Not suitable for use in explosion-endangered areas.
EN / IEC 61326-1,
Class B
approx. 13.5 kg
Rev. 09 / 07.2020
EBA 270
2300-01
100 - 127 V 1
50 - 60 Hz
125 VA
1.25 A
6 x 15 ml
3
1.2 kg/dm
4000
2254
250 Nm
no
Indoors only
2°C to 40°C
2
FCC Class B
51 dB(A)
326 mm
389 mm
239 mm
EN
25/70