Technical Information
Water Cooling Kit Installation
1
Figure 26
2
Cooling may be carried out either through an open circuit with
eventual discharge of the water, or using a closed circuit cooling
system.
The water temperature must be between +15 °C and +35 °C, with an
inlet pressure between 2 and 4 bar.
The water electrical conductance must be ≤ 500 µs/cm. When the conductance
NOTE
is higher, in closed water circuit, the use of up to 20 % of Ethyl-Glycole is
suggested.
212/228
Connect the plate to the pump bottom with four screws M3x20
Assemble the metal or the plastic kit as shown.
Turbo-V 81-T User Manual / 87-900-984-01