Appendix
Guidance and manufacturer's declaration – electromagnetic emission
The Bluephase Style M8 is intended for use in the electromagnetic environment specified below. The customer or the user of the Bluephase Style M8
should assure that it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Bluephase Style M8 is intended for use in the electromagnetic environment specified below. The customer or the user of the Bluephase Style M8
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
± 6 kV contact
Electrostatic
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power supply lines
transient/burst
± 1 kV for input/output lines
IEC 61000-4-4
Surge
± 1 kV line - line
IEC 61000-4-5
± 2 kV line - earth
Voltage dips,
<5% U
T
short interrup-
(>95% dip in U
tions and voltage
40% U
T
variations on
(60% dip in U
power supply
70% U
input lines
T
(30% dip in U
IEC 61000-4-11
<5% U
T
(>95% dip in U
Power frequency
3 A/m
(50/60 Hz)
magnetic Field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
74
Compliance
Electromagnetic environment - guidance
Group 1
The Bluephase Style M8 uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic equipment.
Class B
The Bluephase Style M8 is suitable for use in all establishments, including domestic establish-
ments and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Class A
N/A
Complies
N/A
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line - line
no prot. earth
<5 % U
T
) for 0.5 cycle
(>95% dip in U
T
40% U
T
) for 5 cycles
(60% dip in U
T
70% U
T
) for 25 cycles
(30% dip in U
T
<5% U
T
) for 5 sec
(>95% dip in U
T
30 A/m
Electromagnetic environment - guidance
Floors should be concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at
least 30%.
Mains power quality should be that of typical commercial
or dental environment.
Mains power quality should be that of typical commercial
or dental environment.
Mains power quality should be that of typical commercial
) for 0.5 cycle
or dental environment. If the user of the Bluephase Style M8
T
requires continued operation during power mains interrup-
tions, it is recommended that the Bluephase Style M8 be
) for 5 cycles
T
powered from an uninterruptible power supply or battery.
) for 25 cycles
T
) for 5 sec
T
Power frequency magnetic fields should be at levels charac-
teristic of a typical location in a typical commercial or dental
environment.