Lead Extension - Boston Scientific SC-3138 Série Mode D'emploi

Masquer les pouces Voir aussi pour SC-3138 Série:
Table des Matières

Publicité

Description
The surgical paddle leads function as a component of Precision™, Precision with MultiWave Technology,
Precision Novi™, and Precision Spectra™ Spinal Cord Stimulation systems by delivering electrical
stimulation to the nerve structures in the dorsal aspect of the spinal cord, resulting in an inhibition of pain
sensation.
CAUTION: The Precision SCS System with MultiWave Technology may only be used with surgical
paddle leads if operated at frequencies of 1,200Hz or below.
Surgical Paddle Lead
2x8 Surgical Paddle Lead
The 2x8 Surgical Lead is available in lengths of 50 cm and 70 cm. The distal (paddle) end of the lead has
two columns of eight evenly spaced planar electrodes. Each electrode is 3x2 mm
in size. On the proximal side, this lead employs 2 lead tails. The end of each tail has eight evenly spaced
contacts; identical to the percutaneous leads. The right tail of the paddle lead is laser-etched to allow for
ease of right and left identification. Each tail can be inserted into an IPG, or into a lead extension.
Note: The 2x8 Surgical Lead is intended for use with the Precision, Precision with MultiWave Technology,
Precision Novi, and Precision Spectra SCS systems.
4x8 Surgical Paddle Lead
Two versions of the 4x8 Surgical Lead are available, the CoverEdge 32 Surgical Lead with tightly-spaced
electrodes and the CoverEdge 32 X Surgical Lead with widely-spaced electrodes. Both versions of the
4x8 Surgical Lead are available in lengths of 50cm and 70cm. The distal (paddle) end of the lead has four
columns of 8 evenly spaced planar electrodes. Each electrode is 1x3.4 mm in size. On the proximal side,
the lead employs 4 tails. The end of each tail has 8 evenly spaced contacts; identical to the percutaneous
leads. Each tail of the paddle lead is uniquely identified with marker bands. Reading from left to right with
the electrodes faced down, the first tail has one marker band, the second has two, the third has three,
and the fourth has four marker bands. Each tail corresponds to the respective column of contacts on the
paddle. Each tail can be inserted into an IPG, or into a lead extension.
Note: The 4x8 Surgical Lead is intended for use with the Precision Spectra System only. The 4x8 Surgical
Lead is not compatible with the Precision SCS system, Precision with MultiWave Technology SCS
system or the Precision Novi System.

Lead Extension

Lead Extensions are designed to connect the percutaneous and paddle leads to the Precision, Precision
with MultiWave Technology, Precision Novi, or Precision Spectra IPG for spinal cord stimulation. The
extension may be added to a lead to externalize the lead for a trial procedure or to extend the lead when a
permanent IPG is implanted.
Lead extensions are available in lengths of 25 cm, 35 cm, and 55 cm. The extension can be connected to
either the Trial Stimulator (via an OR Cable Assembly included in the Leads Kit) or directly to an IPG.
Surgical Leads Directions for Use
90970892-03 Rev A 1 of 277

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sc-8120 sérieSc-8216 sérieSc 3138 sérieSc-8336-serieSc-8352-serieSc-8436-serie

Table des Matières