Sterilization
The tip cone assy of a single channel pipette can be repeatedly autoclaved at 121°C (252°F)
(2 ata) for 20 minutes. All other parts and the multichannel pipettes are not autoclavable. After
autoclaving the tip cone assy must be cooled to room temperature for at least two hours. Before
pipetting, make sure that the tip cone assy is dry. We recommend that you check the calibration
after every sterilization cycle.
Autoclavable parts:
Part number 42: single channel tip cone assy
Excluding part number 10 and 14 on 0.5-5ml and 1-10ml models
The part numbers are presented in section Spare Parts.
The table below lists possible problems and their solutions.
Defect
Leakage
Inaccurate
dispensing
Inaccurate
dispensing with
certain liquids
Dispensing knob
jammed to semi
down position
CAUTION!
The Finnpipette is designed to allow easy in-lab service. If you would prefer to
have us or your local representative service your pipette, please make sure that
the pipette has been decontaminated before you send it to us.
Please note that the postal authorities in your country may prohibit or restrict the
shipment of contaminated material by mail.

Trouble shooting

Possible reason
Tip incorrectly attached
Foreign particlesbetween tip and
tip cone
Foreign particles between the
piston, the O-ring and the cylinder
Insuffi cient amount of grease on
cylinder and O-ring
O-ring damaged
Incorrect operation
Tip incorrectly attached
Calibration altered: caused by
misuse, for example
Tip cone (Single channel) or
module (Multichannel) loose
Unsuitable calibration
High viscosity liquids may require
recalibration
One sided force applied to the knob
when adjusting the volume with one
hand
Solution
Attach fi rmly
Clean tip cones attach new tips
Clean and grease O-ring and
cylinder.
Grease accordingly
Change the O-ring
Follow instructions carefully
Attach fi rmly
Recalibrate according to
instructions
Tighten the tip cone or module
with the service tool.
Recalibrate with the liquids
in question
Pull up the knob and press down
again to locked position
14