Digital display

The adjusted delivery volume is clearly indicated in the large
digital display on the handle.
Raw materials
The Finnpipette F1 is made of mechanically durable and chemically resistant materials.
For waste disposal instructions, please contact your local environmental agency. For more
information, please contact us at info.pipettes@thermofi sher.com.
Description of tips
Finntips are recommended for use with the Finnpipette F1.
They are made of virgin natural colour polypropylene, generally regarded as the only
contamination free material suitable for tips. Finntips are also autoclavable (121°C).
The Finnpipette F1 is shipped in a specially designed package containing the following items:
1.
The Finnpipette
2.
Service tool
3.
Multichannel service tool
4.
Finntip sample
5.
Tube of grease (Order No. 2203130)
6.
Instruction manual
7.
Calibration certifi cate / Warranty certifi cate
8.
Shelf hanger (Order No. 2206040)
9.
Two stickers
Cautions are marked with this symbol
Intended use
The intended use of the device is to transfer liquids in the volume range of 0,2 μl to 10 ml.
The F1 Finnpipette pipette and Finntips are designed as a component of an analyzing system
for an end user, who is responsible for validating the system to ensure reliable and safe results.
Damage to health
Follow general procedures for hazard prevention and safety instructions; e.g. wear
protective clothing, eye protection and gloves.
For use and waste disposal of hazardous (e.g. radioactive and potentially infectious)
material, follow the safety instructions and general laboratory practice.
The pipette and tips are not intended for in vivo use. Do not use the pipette for pipetting any
liquid to be injected into a human body.
Do not eject the tip towards anybody.
The Finnpipette F1 is to be used by trained personnel with required laboratory skills. The
instructions for use must be read prior to and during the use of the device (pipette and tip). The
F1 pipette can be used between +4°C and +40°C.
Incorrect dispensing results
Performance may vary due to:
a. pipetting method (forward pipetting technique recommended)
b. temperature (air, liquid, vessel, pipette, and tip)
c. pressure
d. humidity
e. operator, e.g. thumb movement, pipetting angle
f. liquid density, viscosity and vapor pressure
g. type of tip