Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Finnpipette F1 Variable Volume Multichannel
Range
Channels
1.0-10 μl
8, 12, 16
5-50 μl
8, 12, 16
10-100 μl
8, 12
30-300 μl
8, 12

Adjustment

Adjustment is done only for one volume. The recommended adjustment volume is the minimum
volume or 10% of the maximum volume. Multichannel pipettes adjustment check prior
calibration is done using one of the middle channels.
1. Place the service tool into the openings of the
calibration nut at the top of the handle.
2. Turn the service tool clockwise to increase, or
counterclockwise to decrease the volume.
3. After adjustment check the calibration according
to the instructions above.
Formulas for calculating results
Conversion of mass to volume
V = (w + e) x Z
Evaporation loss can be signifi cant with low volumes. To determine mass loss, dispense water
to the weighing vessel, note the reading and start a stopwatch. See how much the reading
decreases during 30 seconds (e.g. 6 mg = 0.2 mg/s).
Compare this to the pipetting time from taring to reading. Typically pipetting time might be 10
seconds and the mass loss is 2 mg (10 s x 0.2 mg/s) in this example. If an evaporation trap or lid
on the vessel is used the correction of evaporation is usually unnecessary.
The factor Z is for converting the weight of the water to volume at test temperature and pressure.
A typical value is 1.0032 μl/mg at 22°C and 95 kPa. See the conversion table on page 62.
Inaccuracy (systematic error)
Inaccuracy is the difference between the dispensed volume and the selected volume of a pipette.
A = V - V
A
=
0
V
=
V
=
0
Inaccuracy can be expressed as a relative value: A% = 100% x A / V
Imprecision (random error)
Imprecision refers to the repeatability of the pipettings. It is expressed as standard deviation (s)
or coeffi cient of variation (cv)
Standard deviation can be expressed as a relative value (CV) CV = 100% x S / V
Volume
Inaccuracy
μl
μl
10
±0,240
1
±0,120
50
±0,75
5
±0,25
100
±1,30
10
±0,50
300
±3,0
30
±1,5
V = volume (μl)
w = weight (mg)
e = evaporation loss (mg)
Z = conversion factor for μl/mg conversion
inaccuracy
mean volume
nominal volume
s
=
standards deviation
v
=
mean volume
n
=
number of measurements
Imprecision Factory
%
s.d. μl
±2,40
0,160
±12,00
0,080
±1,50
0,35
±5,00
0,10
±1,30
0,50
±5,00
0,20
±1,00
0,9
±5,00
0,6
0
11
cv% calibration tip
1,60
Flex 10
8,00
0,70
Flex 200
2,00
0,50
Flex 200
2,00
0,30
Flex 300
2,00
+

Publicité

Table des Matières
loading

Table des Matières