PACKAGE

The FINNPIPETTE is shipped in a
custom made, cardboard package. The
package contains the following items:
1. The pipette
2. Service tool
3. Sample of Finntip
4. Tube grease
5. Instruction Manual

TROUBLE SHOOTING

The accompanying table is a guide to
possible problems and their solutions.
Defect
Leakage
Inaccurate
dispensing
Inaccurate
dispensing
with
certain
liquids
Possible
reason
Tip incorrectly
attached
Foreign
bodies
between tip
and tip cone
Foreign bodies
between the
piston, the O-
ring and the
cylinder
Insufficient
amount of
grease on
cylinder and
O-ring
O-ring
damaged
Incorrect
operation
Tip incorrectly
attached
Calibration
altered:
caused e.g.
by misuse
Unsuitable
calibration.
High viscosity
liquids may
require
recalibration.
-11-
ENGLISH
Solution
Attach firmly
Clean tip
cones attach
new tips
Clean and
grease O-ring
and cylinder.
Use grease
Grease
accordingly
Change the
O-ring
Follow
instructions
carefully
Attach firmly
Recalibrate
according to
instructions
Recalibrate
with the
liquids in
question.