10. Push the tip cone module back to the handle while holding the ejector down, until you
hear a "click".
Service Instructions for Multichannel Pipette Tip Cones
To ensure even performance between all channels in a multichannel pipette, all tip cones
have to be changed at the same time, if any of them needs to be changed. Don't mix tip
cones of different packages, because one bag contains a matched set of tip cones.
Sterilization
The entire pipette can be sterilized by autoclaving it at 121°C (252°F) (minimum 20 minutes).
No special preparations are needed for autoclaving. You can use steam sterilization bags if
needed.
After autoclaving the pipette must be cooled to room temperature for at least two hours.
Before pipetting, make sure that the pipette is dry. We recommend that you check the
calibration after every sterilization cycle to achieve the best possible accuracy.
The table below lists possible problems and their solutions.
Defect
Possible reason
Leakage
Tip incorrectly attached
Foreign particles between tip
and tip cone
Foreign particles between the
piston, the O-ring and the cylinder
Insufficient amount of grease
on cylinder and O-ring
O-ring damaged
Inaccurate
Incorrect operation
dispensing
Tip incorrectly attached
Calibration altered: caused
by misuse, for example
Loosened screws
Unsuitable calibration
Inaccurate
High viscosity liquids may
dispensing with
require recalibration
certain liquids
The Finnpipette FOCUS Multichannel Plus is shipped in a specially designed package
containing the following items:
1.
2.
3.
4.
5.
6.
7.
CAUTION!
The Finnpipette is designed to allow easy in-lab service. If you would
prefer to have us or your local representative service your pipette,
please make sure that the pipette has been decontaminated before you
send it to us.
Please note that the postal authorities in your country may prohibit or
restrict the shipment of contaminated material by mail.

Trouble shooting

Package

The Finnpipette
Service tool
Finntip sample
Tube of grease
Instruction manual
Calibration certificate
Shelf hanger
Solution
Attach firmly
Clean tip cones attach new tips
Clean and grease O-ring and
cylinder.
Grease accordingly
Change the O-ring
Follow instructions carefully
Attach firmly
Recalibrate according
to instructions
Tighten the screws in module
Recalibrate with the liquids
in question
9