Thermo Scientific Finnpipette F1 Guide D'utilisation page 9

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

ISO8655 calibration specifi cations
Thermo Scientifi c Finnpipette F1 specifi cations according to ISO8655 standard.
Finnpipette F1 Variable Volume Single Channel
Range
Volume
0,2-2 μl
0,5-5 μl
1-10 μl
2-20 μl
5-50 μl
10-100 μl
20-200 μl
30-300 μl
100-1000 μl
0,5-5 ml
1-10 ml
Finnpipette F1 Fixed Volume Single Channel
Fixed Volume
μl
10
20
25
50
100
200
250
500
1000
2000
3000
5000
10000
Finnpipette F1 Variable Volume Multichannel
Range
Channel
1-10 μl
8, 12, 16
5-50 μl
8, 12, 16
10-100 μl
8, 12
30-300 μl
8, 12
Inaccuracy
μl
μl
2
±0.080
0.2
±0.080
5
±0.125
0.5
±0.125
10
±0.120
1
±0.120
20
±0.20
2
±0.20
50
±0.50
5
±0.50
100
±0.80
10
±0.80
200
±1.60
20
±1.60
300
±4.0
30
±4.0
1000
±8.0
100
±8.0
5000
±40.0
500
±40.0
10000
±60.0
1000
±60.0
Inaccuracy
μl
%
±0.120
±1.20
±0.20
±1.00
±0.50
±2.00
±0.50
±1.00
±0.80
±0.80
±1.60
±0.80
±4.00
±1.60
±4.00
±0.80
±8.00
±0.80
±16.0
±0.80
±40.0
±1.33
±40.0
±0.80
±60.0
±0.60
Volume
μl
10
1
50
5
100
10
300
30
%
s.d. μl
±4
0.040
±40
0.040
±2.5
0.075
±25
0.075
±1.2
0.080
±12
0.080
±1.0
0.10
±10.0
0.10
±1.0
0.20
±10
0.20
±0.8
0.30
±8.0
0.30
±0.8
0.60
±8.0
0.60
±1.3
1.5
±13
1.5
±0.8
3.0
±8.0
3.0
±0.8
15.0
±8.0
15.0
±0.6
30.0
±6.0
30.0
Imprecision
s.d.μl
0.080
0.10
0.20
0.20
0.30
0.60
1.50
1.50
3.00
6.00
15.00
15.00
30.00
Inaccuracy
μl
%
±0.24
±2.4
±0.24
±24
±1.0
±2.0
±1.0
±20
±0.80
±0.8
±0.80
±8.0
±8.0
±2.7
±8.0
±26.7
9
Imprecision
cv%
2.0
20.0
1.5
15
0.8
8.0
0.5
5.0
0.4
4.0
0.3
3.0
0.3
3.0
0.5
5.0
0.3
3.0
0.3
3.0
0.3
3.0
cv%
0.80
0.50
0.80
0.40
0.30
0.30
0.60
0.30
0.30
0.30
0.50
0.30
0.30
Imprecision
s.d.μl
cv%
0.16
1.6
0.16
16
0.4
0.8
0.4
8.0
0.30
0.3
0.30
3.0
3.0
1.0
3.0
10.0

Publicité

Table des Matières
loading