Motor not running (continued)
7. Check capacitor
- Correctly connected?
ok?
8. Check electronic module
- Correctly connected
- Defective
- Check the jumper setting
ok?
9. Mount drive unit
10. Close device
11. Functional test
Xll
Jumper settings
• Check the jumper setting. Also see the jumper settings (foldout)
• Check correct position of the On/Off switch
c / i - v e r s i o n s
On
8 sec.
16 sec.
*
32 sec.
8
EN
not OK
not OK
Off
4 sec.
8 sec.
16 sec.
Connect cord
Replace capacitor
Connect cable properly
Set jumper correctly
Replace defective
electronic module
a. secure drive unit with 4 screws
b. AC/DC version: mount battery
b. mount QuatroFlex™
b. Mount sound absorber