Guidance and manufacturer's declaration – electromagnetic immunity
This breast pump is intended for use in the electromagnetic environment specified below. The customer
or the user of the breast pump should assure that it is used in such an environment.
IEC 60601
Immunity test
test level
Conducted RF
3 V
IEC 61000-4-6
0,15 MHz – 80 MHz
6 V in ISM and
amateur radio bands
between 0,15 MHz
and 80 MHz
80% AM at 1 kHz
10 V/m (minimum)
Radiated RF
80 MHz – 2,7 GHz
IEC 61000-4-3
80 % AM at 1 kHz
Questions? Visit www.medela.com or call us at 1-800-435-8316.
Compliance
Electromagnetic
level
environment – guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the breast pump, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
6V
transmitter.
Recommend separation distance
d
= 1,2
d
= 1,2√P 80 MHz - 800 MHz
d
= 2,3√P 800 MHz - 2.7 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
10 V/m
distance in metres (m).
(minimum)
Field strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey,
compliance level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol.
P
a
should be less than the
51
b