MSA Solaris Manuel D'exploitation page 50

Détecteur multigaz
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Table 6-11. Nitrogen (Appropriate Models Only) Dioxide Typical
Performance Specifications
RANGE
RESOLUTION
REPRODUCIBILITY
RESPONSE TIME
*See TABLE 6-2 NOTE.
Table 6-12. Nitrogen Dioxide (Appropriate Models Only) Cross
Reference Factors for Solaris Calibration
NOTE: Data is presented as the indicated output in ppm, which would result from the
application of 10 ppm of the test gas
TEST GAS (10 PPM)
Hydrogen Sulfide (H 2 S)
Sulfur Dioxide (SO 2 )
Carbon Monoxide (CO)
Nitric Oxide (NO)
Ammonia (NH 3 )
Methane (CH 4 )
6-8
50.0 ppm
0.1 ppm H 2 S, for 0.5 to 50.0 ppm NO 2
+0.5 ppm NO 2 or 10% of reading, whichever
is greater (normal temperature range*)
+0.5 ppm NO 2 or 20% of reading, whichever
is greater (extended temperature range*)
90% of final reading in 60 seconds
(normal temperature range*)
EQUIVALENT PPM
-12.7 +1.2
-.6 +.1
0 +0
0 +0
0 +0
0 +0

Publicité