Table 6-7.
CARBON MONOXIDE (appropriate models only) -
Typical Performance Specifications
RANGE
500 ppm CO
RESOLUTION
1 ppm CO, for 5 to 500 ppm CO
REPRODUCIBILITY
+5 ppm CO or 10% of reading, whichever is greater
0 to 300 ppm CO, +15% >300 ppm CO
(normal temperature range*)
+10 ppm CO or 20% of reading, whichever is greater
(extended temperature range*)
RESPONSE TIME
90% of final reading in 60 seconds (normal temperature range*)
*See TABLE 6-2 NOTE
Table 6-8. CARBON MONOXIDE - Cross Reference Factors
for Solaris Calibration Using Calibration Cylinder (P/N 10045035)
NOTE:
Data is presented as the indicated output in ppm, which would result from the
application of 100 ppm of the test gas.
TEST GAS (100 PPM)
Carbon Monoxide (CO)
Hydrogen Sulfide (H 2 S)
TEST GAS (100 PPM)
Sulfur Dioxide (SO 2 )
Nitrogen Dioxide (NO 2 )
Nitric Oxide (NO)
Chlorine (Cl 2 )
Ammonia (NH 3 )
Hydrogen Chloride (HCl)
Ethylene (C 2 H 4 )
Hydrogen Cyanide (HCN)
Methane (CH 4 )
Ethanol (EtOH)
Hydrogen (H 2 )
6-6
EQUIVALENT PPM
100 +9
4 +4
EQUIVALENT PPM
0 +1
2 +6
70 +10
1 +8
2 +4
3 +2
90 +9
0 +1
0 +0
4 +5
70 +26