Configuration Via Display And Operating Unit; Configuring Pressure Measurement; Configuring Level Measurement - WIKA DMSU21SA Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

8. Configuration via display and operating unit

8. Configuration via display and operating unit

8.1 Configuring pressure measurement

EN
1. Open the operating menu with [↵].
Select "Basic setting" and confirm with [↵].
2. Select "Application" and confirm with [↵].
3. Select "Pressure" and confirm with [↵].
4. Select "Unit" and confirm with [↵].
5. Select pressure unit and confirm with [↵].
Pressure unit is set.
6. Go back one menu level using [ESC].
Select "Mode" and confirm with [↵].
7. Select "Pressure" and confirm with [↵].
» Mode is set.
8. Scale the measuring range.
→ See chapter 8.6 "Scaling the measuring range".
9. Perform a mounting correction.
→ See chapter 8.8 "Mounting correction (offset)".
» Pressure measurement is configured.

8.2 Configuring level measurement

Requirement
1. Open the operating menu with [↵].
Select "Basic setting" and confirm with [↵].
2. Select "Application" and confirm with [↵].
28
Length unit for the fill level is known.
The density of the medium is known
WIKA operating instructions, model DMSU21A

Publicité

Table des Matières
loading

Table des Matières