8. Configuration ... / 9. Diagnostic functions
8.12 Change PIN
Factory setting: 0000
1. Open the operating menu with [↵].
EN
Select "Basic setting" and confirm with [↵].
2. Select "Write protect" and confirm with [↵].
3. Select "change PIN" and confirm with [↵].
4. Change digit using [▲] [▼] and confirm with [↵]. The cursor
moves to the next digit. Repeat for each digit.
» Pin is changed.
9. Diagnostic functions
9.1 Performing a pressure simulation
A pressure value within the measuring range must be entered and is converted into a
current value and output.
1. Open the operating menu with [↵].
Select "Diagnostic" and confirm with [↵].
2. Select "Simulation" and confirm with [↵].
3. Select "Press. simu." and confirm with [↵].
4. Change digit using [▲] [▼] and confirm with [↵]. The cursor
moves to the next digit. Repeat for each digit.
» Simulation is active.
5. Ending the simulation. Press [ESC] to do this.
42
WIKA operating instructions, model DMSU21A