5. Commissioning, operation
5.4 Commissioning
During the commissioning process pressure surges must be avoided at all costs. Open
the shut-off valves slowly.
EN
5.5 Electrical mounting
DANGER!
Danger to life caused by electric current
Upon contact with live parts, there is a direct danger to life.
▶
The instrument may only be installed and mounted by skilled electrical
personnel.
▶
Operation using a defective power supply unit (e.g. short-circuit from the
mains voltage to the output voltage) can result in life-threatening voltages
at the instrument!
5.5.1
Requirements for hazardous areas
Power the process transmitter via an intrinsically safe circuit (Ex ia). Both the
■
effective internal capacitance and inductance must be considered (→ See chapter 15
"Specifications").
Provide the required voltage supply separation between Ex and non-Ex areas with a
■
certified isolated barrier or Zener barrier (suitable isolated barrier, model IS Barrier).
- For applications that require EPL Gb, the power supply and signal circuit should
■
have a protection level of "ib". Then the interconnections and the transmitter will have
a protection level of II 2G Ex ib IIC T4/T5/T6 Gb, even if the process transmitter is
marked otherwise (→ See EN 60079-14 section 5.4).
5.5.2
Requirements for connection cable
Use and assemble connection cable that is suitable for the application. For cables with
■
flexible wires, always use ferrules appropriate for the wire cross-section.
Where there is electromagnetic radiation above the test values per EN 61326, a
■
shielded connection cable must be used.
When using an M12 x 1 (4-pin) circular connector, the mating connector is supplied by
■
the customer. Ensure the matching design from the connector manufacturer.
Electrical connections
Spring-loaded terminals
Cable gland M20 x 1.5, stainless
steel in hygienic design
Angular connector DIN 175301-803A
with mating connector
Circular connector M12 x 1 (4-pin)
without mating connector
20
Wire cross-section:
Wire or strand: 0.2 ... 2.5 mm² (AWG 24 ... 14)
Strand with end splice: 0.2 ... 1.5 mm² (AWG 24 ... 16)
Cable diameter: 6 ... 12 mm [0.24 ... 0.47 in]
Cable diameter: 6 ... 8 mm [0.24 ... 0.31 in]
Wire cross-section: max. 1.5 mm² (AWG 16)
Observe manufacturer's specifications
WIKA operating instructions, model DMSU21A