Connection to a surgical aspirator
Connect the mask circuit gas exhaust tube to the supplied fitting (Figura 5 Pos 17).
Connect the fitting to the surgical aspirator tube (Ø11 mm).
Valve adjustment
Turn the perforated ring nut on the valve (Figura 5 Pos 17) until the holes close.
Start gas delivery by setting ~10 l/min.
Start the surgical aspirator by setting the minimum aspiration value.
Observe the bar (Figura 8 Pos F).
Turn the perforated ring nut on the valve gradually opening the holes until the bag reaches
its ideal condition of "HALF FULL HALF EMPTY"
 
ATTENTION
By changing the flow delivery it may be necessary to adjust the correct aspiration
 
flows of the exhaled gases.
16