LEVEL 1:
n Extension (stretching)
− Maximum knee extension:
− Maximum hip extension:
n Flexion (bending)
− Maximum knee flexion:
− Maximum hip flexion:
Note!
The programmed value and the value
measured at the patient's knee may
differ slightly.
n Speed
The speed can be adjusted in steps of 5%
between 5% and 100%.
Default setting: 50 %
n Warm-up protocol
During warm-up, the patient will slowly
become used to the set maximum ex-
tension and flexion values, starting from
the center position.
The warm-up protocol starts in the mid-
dle between the two maximum values
set for stretching and bending. The range
of motion increases with each cycle, until
the programmed maximum values are
reached after a total of 15 cycles.
Default setting: disabled
LEVEL 2:
n Extension pause
Pauses occur at the set extension limit,
just before the bending movement
starts. Pauses can be set to any value
between 0 and 59 seconds in steps of 1
second, and to values between 1 and 59
minutes in steps of 1 minute.
Default setting: no pause
n Flexion pause
Pauses occur at the set flexion limit, just
before the stretching movement starts.
Pauses can be set to any value between
0 and 59 seconds in steps of 1 second,
− 10°
and to values between 1 and 59 minutes
7°
in steps of 1 minute.
Default setting: no pause
120°
n Timer (therapy time)
115°
Default setting is continuous operation.
A clock symbol in the upper right-hand
corner of the display identifies the con-
tinuous mode of operation. The clock
indicates the elapsed therapy time. In
the continuous mode, the device must
be stopped with the STOP button.
However, you can also select therapy
durations of 1 to 59 minutes in steps of
1 minute and of 1 to 24 hours in steps of
30 minutes.
When the time has elapsed, the unit
switches automatically off and stops
in the position equivalent to the set
extension value +10°. In this case, a circle
replaces the clock symbol. The circle fills
as the therapy time progresses.
n Reverse on load
(safety function)
The device automatically starts moving in
the opposite direction of the last move-
ment when the patient's resistance (load)
exceeds the set value. Adjustable levels
for reverse on load feature: 1 - 25. At
level 1, very low resistance will cause the
device to reverse; at level 25, a high resis-
tance is required to initiate the reversal.
Default setting: Level 25
Caution!
m
Patient hazard –
The reverse-on-load feature is only a
safety measure in the event of cramps,
spasms, locked joints and similar condi-
tions. The manufacturer cannot be held
liable for misuse of this feature.
51