8.3 Description of individual menu items
8.3.1 Weighing Units
 In weighing mode keep the PRINT key pressed until [Unit]
 Press SET key, the selected unit is displayed.
 Use the MODE key to select between the different units (see
 Use the SET key to confirm the selected unit
Gram
Ounce
Troy ounce
Tael Hongkong
Tael Taiwan
Grain (dependent on model)
Pennyweight
(dependent on model)
Momme
Tola
Carat (dependent on model)
Freely selectable factor *)
*) Input conversion factor
 As specified above, press repeatedly the MODE key until „FFA"
 To enter the factor, press the SET key; the enabled digit starts
 Confirm input by pressing the SET key.
 Press repeatedly the SET key to take over the „Freely
30
appears.
following table).
appears.
flashing.
Using the MODE button, the displayed value is increased by 1,
with the PRINT button it is reduced by 1.
With the TARE key selection of the number to the left.
selectable factor" as current weighing unit.
Display
Conversion factor *
1 g =
g
1.
oz
0.035273962
ozt
0.032150747
tlh
0.02671725
tlt
0.0266666
gn
15.43235835
dwt
0.643014931
(mom)
0.2667
tol
0.0857333381
ct
5
FFA
xx.xx