Télécharger Imprimer la page

KERN PCD 300-3 Mode D'emploi page 69

Masquer les pouces Voir aussi pour PCD 300-3:

Publicité

Les langues disponibles

Les langues disponibles

7.5 Pre-Tare function
M O D E
C A L
Delete PRE-TARE
value
M O D E
C A L
7.6 Plus/minus weighings
TA R E
PCD-BA-e-1615
Using this function the weight of a tare vessel is stored. Even
after turning off/on the weighing balance will continue working
with the saved tare value.
 In weighing mode put tare vessel on the weighing plate
 Press repeatedly the MODE key until „PtArE" flashing
appears.
 Use SET key to store the current weight on the weighing
plate as a PRE-TARE value.
 Unload the balance and set the TARE button to zero
 Press repeatedly the MODE key until „PtArE" flashing
appears.
 Use the SET key to confirm. The PRE-TARE value is
deleted, the zero display appears.
For example unit weight control, fabrication control etc.
 Put the nominal weight on the weighing plate and tare
using the TARE button.
 Remove the nominal weight
 Put the test objects subsequently on the weighing plate,
the respective deviation from the nominal weight is
displayed with the respective sign to „+" and „-".
According to the same procedure also packages with the
same weight can be produced, referring to a nominal
weight.
 Back to weighing mode by pressing the TARE button.
21