 Place fifth adjustment weight and acknowledge
by
. The scales will change to zero
display.
 Take away adjustment weight. After a short
time CAL F is displayed.
Finally the balance will switch off automatically. Now the linearization is concluded
successfully.
In case of an adjustment error or incorrect adjusting weight the display will show an
error message; repeat linearization process.
5.8 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory).
This adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
16
PCD-BA-e-1615