8.3.4 Filterfunction
(except PCD 10K-3)
(Example)
PCD-BA-e-1615
This menu item allows the balance to be set according to
specific ambient conditions and measuring purposes.
 In weighing mode keep the PRINT key pressed until [Unit]
appears.
 Press the MODE button several times until „StAbiL" is
displayed.
 Acknowledge using SET button, the current setting is
displayed.
 Select the desired settings by pressing the MODE button.
Filter 1:
1
The balance reacts quickly and in a sensitive manner, quiet set-
up location.
Filter 2:
2
The scale reacts normally, normal installation site
Filter 3:
3
The balance reacts slowly and in a robust manner, busy set-up
location
 Use the SET key to confirm selection.