8.3.6 Animal weighing function
The animal weighing function can be applied for busy weighings.
During a defined period the mean value of the weighing results is
formed.
The more unquiet the weighed item, the longer the period should
be selected.
In weighing mode keep the PRINT key pressed until "Unit"
Press the MODE button several times until „ANL" is displayed.
Acknowledge using SET key, the current setting is displayed.
Select the desired settings by pressing the MODE key
Use the SET key to confirm selection.
Put the weighing good (animal) on the weighing plate and
Use the SET key to change between animal weighing and
Press the SET key to restart the weighing cycle for animal
PCD-BA-e-1615
appears.
Period 3 seconds
ANL
3
Period 5 seconds
ANL
5
ANL
10
Period 10 seconds
ANL
15
Period 15 seconds
ANL
off
Animal weighing not active
press the SET button. In the display runs a „Countdown".
The average value of the weighing results is displayed and
remains displayed on the screen.
normal weighing.
weighing.
35