Storage conditions
4.2
Fastening the transport lock
1
3
2
1
Motor cover
2
Edge of the centrifuging
chamber
3
Bellows
AB1401en
■
Before transporting, fasten the transport lock and disconnect the device
from the mains socket.
■
The transport temperature must be between -20 °C and +60 °C.
■
Humidity must not be condensing. Humidity must be between 10% and
80%.
■
Be aware of the weight of the device.
■
When transporting using a transport aid (e.g., a pallet truck), the trans-
port aid must be able to carry at least 1.6 times the transport weight of
the device.
■
Secure the device to prevent it tipping over and falling down during
transport.
■
Never transport the device sideways or upside down.
■
The device must be stored in the original packaging.
■
Only store the device in dry rooms.
■
The storage temperature must be between -20 °C and +60 °C.
■
Humidity must not be condensing. Humidity must be between 10% and
80%.
Personnel:
■
Trained user
1.
Open the lid.
2.
For UNIVERSAL 320 R:
Check the bellows (
seating.
3 ) must be placed over the edge of the motor cover ( 1 )
The bellows (
and over the edge of the centrifuging chamber (
3.
Close the lid.
3
1
3
3
1
3
Fig. 13: Transport lock
1
Transport lock
2
Screw
3
Spacer sleeve
4.
Lay the device down on its right-hand side.
Rev.: 13 / 11.2023
Transport and storage
3 ) underneath the motor cover for correct
3
1
3
3
2 ).
2
1
17 / 44