9.3
Emergency release
Fig. 22: Emergency release
1
Hole
10
Disposal
10.1
General instructions
AB1401en
2.
Wait 10 seconds.
3.
Set the mains switch to
The lid cannot be unlocked by the motor in the event of a power failure.
Emergency unlocking by hand must be performed.
Risk of electric shock due to maintenance and servicing
work on live device.
−
WARNING
Danger of cutting and crushing due to moving rotor.
−
Personnel:
■
Trained user
1.
Look through the window in the lid to ensure that the rotor is sta-
tionary.
2.
Insert the hex key horizontally into the hole (
until the lid opens.
3.
Remove the hex key from the hole (
1
4.
Check whether the left side of the
power is restored.
When the left side of the
[STOP/OPEN] button so that the motorised lid lock assumes the
home position (open) again.
The device can be disposed of via the manufacturer.
A Return Material Authorisation (RMA) form must always be
requested for a return.
If necessary, contact the Technical Service Department of
the manufacturer:
−
−
−
−
−
Rev.: 13 / 11.2023
[I] .
WARNING
Disconnect the device from the mains before carrying
out repairs and maintenance.
Do not open the lid until the rotor has stopped.
1 ).
[STOP/OPEN] button flashes when
[STOP/OPEN] button flashes, press the
Andreas Hettich GmbH & Co. KG
Föhrenstraße 12
78532 Tuttlingen, Germany
Phone: +49 7461 705 1400
E-Mail: service@hettichlab.com
Disposal
1 ) and turn clockwise
41 / 44