Instructions for Use
Installation

Control Unit Connection

WARNING!
Voltage delivered by control units used with the VacIon pumps are potentially
lethal. To operate the pump safely ensure that grounding connection is
correctly made.
Connect the control unit to the pump as follows:
1
2
42/52
Ground the pump to the power supply by placing the ground
spring over the pump insulator before attaching the high voltage
lead. On cables with ground leads, attach the ground lead to a
ground lug on the system.
The power supply, in turn, is grounded through its power lead.
Therefore, to ensure a completely grounded circuit, the
threeprong plug on the power supply cord must be inserted in a
socket that has a ground connection. Do not use socket
convertors that prevent the completion of this ground circuit.
Miniature and 2 l/s VacIon Pumps User Manual 87-400-044-01 (E)